Skip to content

SOP — Anton Paar DMA 35 Maintenance

FUERST WIACEK Version 1.2 | Lab — Anton Paar DMA 35 portable density meter | Frequency: clean after each session, calibration check monthly


Purpose

This SOP covers the full cleaning procedure, calibration, and maintenance schedule for the Anton Paar DMA 35 portable density meter used for daily Plato readings during fermentation monitoring. Proper maintenance ensures measurement accuracy and extends the life of the instrument.

For daily quick-rinse after use, refer to the note in the Fermentation Monitoring SOP.


Equipment

  • Anton Paar DMA 35
  • Small pitcher
  • Washing-up liquid (Spüli)
  • Fresh water
  • Distilled water (for calibration)
  • Alcohol (isopropanol)

Cleaning Procedure — After Each Use

Perform this full cleaning at the end of each session or at the end of the day.

  1. Prepare a small pitcher of warm water with a single drop of Spüli.
  2. Flush the instrument 2–3 times with the soapy water. Expel after each draw.
  3. Flush 2–3 times with fresh water. Expel after each draw.
  4. Flush once with alcohol. Expel.
  5. Draw air through using the instrument's integrated pump until the measuring cell is visibly dry.

Note

The alcohol flush displaces residual water and speeds drying. Do not skip it.

Warning

Never leave liquid sitting in the instrument when storing. Always complete the air-drying step.


Calibration

The DMA 35 is calibrated using distilled water as a reference. Calibrate whenever readings appear inconsistent, after the instrument has been knocked or dropped, or after any service.

Calibration Procedure

  1. Ensure the instrument is clean and dry — run the full cleaning procedure first.
  2. Open the menu on the DMA 35 and navigate to the calibration function.
  3. Draw a fresh sample of distilled water into the instrument. Follow the on-screen prompt and confirm when stable.
  4. Log the calibration date and result in the maintenance log below.

Note

Use fresh distilled water only. Do not use water that has been sitting open.


Maintenance Schedule

Task Frequency Notes
Full cleaning (soapy water → fresh water → alcohol → air dry) After each session See cleaning procedure above
Calibration check Monthly Log result in maintenance log
Visual inspection (seals, cap, casing) Monthly Check for cracks, wear, or damage
Battery check As needed Replace when low battery indicator appears
Full service / factory check As recommended by Anton Paar Contact Anton Paar or authorised service partner

Maintenance Log

Calibration entries, services, and incidents are recorded in the Anton Paar maintenance log — see Spreadsheets reference for the location.

For each calibration, log:

  • Date
  • Result (pass / fail / value)
  • Notes — anything unusual (knock, drift, replaced parts)
  • Initials

Troubleshooting

Issue Likely Cause Action
Reading inconsistent or drifting Residue in measuring cell Perform full cleaning procedure. Recalibrate.
Reading consistently high or low vs. reference Calibration drift Recalibrate with distilled water.
Instrument slow to stabilise Sample contains bubbles or residue Flush more thoroughly before reading. Ensure no gas pockets.
IPA not available Missing consumable Substitute with a longer air-dry sequence. Restock alcohol.
Display shows error after knock or drop Potential sensor damage Do not use. Contact Anton Paar for inspection.